User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Measurements Group 1550A:Strain Indicator Calibrator DATE: 14-Sep-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 2 NUMBER OF LINES: 56 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P M F W 1.002 ASK+ X 1.003 HEAD INITIAL CONDITIONS 1.004 DISP Connect the 5700A and the UUT as follows. 1.004 DISP 5700A TO 1550A 1.004 DISP OUTPUT V ohm A HI ......... 120 ohm P+ TERMINAL 1.004 DISP OUTPUT V ohm A LO ......... 120 ohm P- TERMINAL 1.004 DISP SENSE V ohm HI ............ 120 ohm P+ TERMINAL 1.004 DISP SENSE V ohm LO ............ 120 ohm P- TERMINAL 1.005 DISP Connect the 34401A and the UUT as follows. 1.005 DISP 34401A TO 1550A 1.005 DISP INPUT HI .................. 120 ohm S+ TERMINAL 1.005 DISP INPUT LO .................. 120 ohm S- TERMINAL 1.006 DISP Set the UUT 120 ohm bridge S+ shorting link to the 1.006 DISP closed position. 1.007 DISP Set the UUT front panel controls as follows: 1.007 DISP MICROSTRAIN buttons ....................... 000 1.007 DISP CIRCUIT button ............................ 120 ohm 1.007 DISP POLARITY button ........................... positive (+) 1.008 HEAD {} 1.009 HEAD {* 120 Ohm Bridge Accuracy Test *} 1.010 CALL Measurements Group 1550A 120 Ohm Sub 1.011 JMP 2.001 1.012 EVAL 2.001 HEAD {} 2.002 HEAD {* 350 Ohm Bridge Accuracy Test *} 2.003 DISP Connect the 5700A and the UUT as follows. 2.003 DISP 5700A TO 1550A 2.003 DISP OUTPUT V ohm A HI ......... 350 ohm P+ TERMINAL 2.003 DISP OUTPUT V ohm A LO ......... 350 ohm P- TERMINAL 2.003 DISP SENSE V ohm HI ............ 350 ohm P+ TERMINAL 2.003 DISP SENSE V ohm LO ............ 350 ohm P- TERMINAL 2.004 DISP Connect the 34401A and the UUT as follows. 2.004 DISP 34401A TO 1550A 2.004 DISP INPUT HI .................. 350 ohm S+ TERMINAL 2.004 DISP INPUT LO .................. 350 ohm S- TERMINAL 2.005 DISP Set the UUT 350 ohm bridge S+ shorting link to the 2.005 DISP closed position. 2.006 DISP Set the UUT front panel controls as follows: 2.006 DISP MICROSTRAIN buttons ....................... 000 2.006 DISP CIRCUIT button ............................ 350 ohm 2.006 DISP POLARITY button ........................... positive (+) 2.007 CALL Measurements Group 1550A 350 Ohm Sub 2.008 HEAD {} 2.009 END